描述

| Operating Principle | Partial pressure electrochemical |
| Output | 9 mV to 13 mV in 210 mBar O2 |
| Measurement Range | 0 mBar to 1500 mBar O2 |
| Response Time | <15 s (air to 100% O2 ) |
| Baseline Offset | <200 μV |
| Linearity | Linear 0% to 100% O2 |
| Temperature Compensation | Partial compensation for Teledyne T-7 equivalence |
| External Load Resistor | 10 kΩ min |
| Connector | UK Data Socket AMP (P/N 520257-2) |
| Housing Material | White ABS |
| Typical Applications | Critical care anesthesia |
| Operating Temperature Range | -20°C to +50°C |
| Operating Pressure Range | 0.5 Bar to 2.0 Bar |
| Operating Humidity Range | 0% to 99% RH noncondensing |
| Long-Term Ouput Drift in 100% O2 | < 5% signal loss/year |
| Recommended Storage Temperature | -10°C to 40°C (short excursions to 50°C allowed) |
| Expected Operating Life: @ 20°C @ 40°C | 1.5 x 106 % O2 hours 0.8 x 106 % O2 hours |



We accept payment via TT (Telegraphic Transfer) and L/C. It’s minimal amount required for L/C. For small orders of
samples, it’s acceptable by Western Union ,PayPal and Money gram



















Reviews
There are no reviews yet.