Envitec OOM201 Medical Oxygen Sensor O2 Sensor
$80.00
Envitec OOM201 Medical Oxygen Sensor O2 Sensor
描述

| Measurement range | 0 % – 100 % oxygen (at atmospheric pressure) |
| Nominal sensor lifetime | ≥ 500 000 % volume oxygen hours |
| Output in ambient air | 14 mV-20.7 mV (Dual Cathode), load 600 Ohms |
| Electrical interface | Gold plated slip rings |
| Accuracy meets | ISO 80601-2-55 requirements |
| Repeatability | < 1 % volume O2 at constant temperature and pressure |
| Linearity error | < 3 % relative |
| Response time | < 12 s to 90 % of final value |
| Zero offset voltage | < 200 µV in 100 % nitrogen, applied for 5 min |
| Cross interference | meets ISO 80601-2-55 requirements |
| Influence of humidity | -0.03 % rel. per % RH at 25 °C |
| Pressure range | 0.6 bar-2 bar (ppO2 0 … 1250 mbar O2) |
| Influence of pressure | proportional to change in oxygen partial pressure |
| Influence of mechanical | shock < 1 % relative after a fall from 1 m |
| Operating temperature | 0 °C ~+50 °C |
| Temperature compensation | no temperature compensation |
| Operating humidity | 0 %- 99 % RH non-condensing |
| Long term output drift | < 1 % volume oxygen per month |
| Typically | < -15 % relative over lifetime |
| Storage temperature | -20 °C~ +50 °C |
| Recommended storage | +5 °C ~ +15 °C |
| Recommended load | ≥ 10 kOhms |
| Warm-up time | < 30 minutes, after replacement of sensor |



We accept payment via TT (Telegraphic Transfer) and L/C. It’s minimal amount required for L/C. For small orders of
samples, it’s acceptable by Western Union ,PayPal and Money gram


















Reviews
There are no reviews yet.